Type:CX-IDO-1301A industrial dissolved oxygen meter

Illustrate:It is widely used for continuous monitoring of the dissolved oxygen content of various solutions used in power plants and biological and chemical industries. And it’s equipped with a microprocessor

CX-IDO-1301A Industrial Dissolved Oxygen Meter


Main features
1.Large vacuum fluorescent display (VFD) and man-machine interaction function.
2.There are application analysis for the electrode and four calibration methods.
3.The functions of automatic temperature compensation and automatic range switching.
4.Available with history curve, real-time trend chart and clock.
5.There are four kinds of built-in recorders. It is available with isolated current output and alarm.
6.The user can set the password or consult the “Help” menu or the “Service Guide” Menu.
7.It has power-off protection. The data can be saved for 5 years under a continuous power-failure condition.
8.The communication interface can be connected directly to a computer and enables remote operation and control.
9.The instrument adopts universal module power supply.
10.It is equipped with the dissolved oxygen electrode with excellent technical performance and stable and reliable
measuring capacity. It can be easily used and maintained.

Main parameters

Measuring range and error 0-20µg/L; 0-200µg/L; 0-20mg/L (automatic range switching).
Resolution: 0.1µg/L; error: ±0.2 µg/L
Resolution: 0.01 mg /L; error: ±0.02 mg/L
Temperature: 0~60C; resolution: 0.1C; error: ±0.8C
Range of automatic temperature compensation 0~60C, with 25C as the reference temperature
Residual signals of the electrode <1‰ (air signal at room temperature)
Response time <60sWhen the temperature is 25C
<30s when the temperature is 35C
Stability Under normal pressure and constant temperature,
the weekly shift should be <2%F.S. 
Water sample pressure 0.6MPa
Isolated output 0-10mA (load resistance <1.5KΩ).4-20mA
(load resistance<750Ω)
Data storage capacity 55 hours (1 point/10 seconds); 14 days (1 point/1 minute) ;
68 days (1 point/5 minutes); 412 days (1 point/30 minutes)
Normally-open contact of the alarm relay 3A 240V AC, 6A 28V DC or 120V AC
Overall dimension of the electrode unit 151 (length) x 151 (width) x 251 (height) mm
Dimension of the mounting hole 138 x 138mm
Power supply 85-265VAC 50±1Hz
Connection pattern of the inlet and outlet water pipes Φ6, Φ8, Φ10, Φl2 and Φl4
Note:More sensor types and installation methods are: flow-through type,duct-type, plug-in type